75 mm
i.d./360 mm
o.d. fused silica/Upchurch Fingertight connectors
Resolution/Lineshape
(1H)
(5% CHCl3
in acetone-d6, stopped flow, flowcell filled, LB=0)
50%
< 1 Hz
0.55%
< 10 Hz
0.11%
< 20 Hz
Proton 90°
Pulse Width
Indirect Detect
Carbon 90°
Pulse Width
£ 5
ms
£
15 ms
VT Control*
(gas source supplied by customer)
0 - 50 °C
Proton Signal to
Noise
10 mM sucrose in
90% D2O/10% ACN-d3, quantity sufficient to
overfill flowcell. Anomeric proton. LB=0.7 Hz.
> 30:1 single scan
RF connectors
BNC
RF Homogeneity (450°/810°)
75%/50%
Gradient Specifications
Strength:
(typical)
35 G/cm/A
Maximum recommended duty cycle
< 10 %
Maximum
recommended drive current
< 10 A
Capillary NMR
Installation Kit
Assortment of
capillary unions, sleeves, ferrules, fittings, 10 m of 50/360
capillary, diamond wheel capillary cutter, and in-line solvent
filter replacement. An important toolkit for startup and periodic
maintenance.
* For Bruker and JEOL
systems, probe interfaces to spectrometer manufacturer’s heater,